You change the oil, send the next sample, and the oxidation number comes back higher. It is an awkward result to explain after a maintenance job that was meant to give the machine a fresh start.
Before treating it as rapid deterioration, find the unused-oil result. If the formulation changed, the new oil may have started above the final reading from the old fill. The report can be accurate while the line joining those two results tells the wrong story.
New oil has a signal of its own
FTIR measures the absorption of infrared light. Base oils and additives contribute to the spectrum before the lubricant enters a machine; service then changes that starting chemistry.
Thermo Fisher's lubricant-analysis note shows four unused oils with different spectra, alongside a new oil, a used sample and their difference spectrum. Its point is practical: the starting formulation can affect the reported number. Matching viscosity grades do not establish matching infrared spectra.
Oxidation also produces a mixture of compounds. An FTIR oxidation result tracks a selected spectral feature associated with that chemistry, rather than counting one substance called oxidation. Spectro Scientific explains the chemistry and measurement.
That is why the number needs its oil identity and measurement basis beside it.
Five results, with a break at the refill
Here is a hypothetical set of direct FTIR readings on one consistent basis: absorbance per 0.1 mm optical path. These are not laboratory results from an actual compressor.
| Oil and sample | Direct reading |
|---|---|
| Unused formulation A | 0.06 |
| Last in-service sample from A | 0.12 |
| Unused formulation B | 0.18 |
| First in-service sample from B | 0.20 |
| Later comparable sample from B | 0.25 |
Across the oil change, the report appears to rise by 0.20 − 0.12 = 0.08. But the unused oils differ by 0.18 − 0.06 = 0.12. The cross-fill step contains a change of formulation as well as a change of service history. It cannot tell you how much deterioration occurred after the refill.
Within B's series, 0.25 − 0.20 = 0.05 is a more useful lead. Provided the samples are comparable, something has changed since B's first in-service sample. Whether that change warrants action depends on the method's precision, the equipment history and the other findings.
Keep all five results on the chart. Leave a visible break where the formulation changes. You lose useful history if you delete A, and create a misleading trend if you join A to B.
Ask the lab which number you are reading
The column heading may say oxidation on every report, even when the measurement approach differs.
ASTM D7414's public scope distinguishes direct readings from measurements made after subtracting a reference-oil spectrum. It also specifies different reporting conventions. Subtracting two printed results yourself does not reproduce that spectral subtraction. A switch between direct and differential reporting therefore needs its own break in the series.
Ask for the method, approach, units and identity of any reference oil. Then ask whether the earlier spectra can be compared on that same basis. Instrument acquisition conditions matter too; ASTM D7418 covers setup and operation intended to support comparable FTIR data.
There is an important scope limit: D7414 excludes ester-based oils. A move to an ester-based compressor lubricant needs a suitable method agreed with the laboratory, rather than an assumption that the old test still applies.
Keep the unused bottle and the first circulating sample
They answer different questions. The unused sample shows the delivered product's starting chemistry. The first representative circulating sample shows the oil actually in the machine, which can include retained oil from lines and coolers or subsequent makeup oil.
Link the unused sample to the product and delivery lot, and link the circulating sample to the fill date and operating hours. Tell the laboratory about the formulation change. A familiar product name in its reference library is little help if it belongs to a different formulation.
The broader trend still needs equipment context. ASTM D7669 addresses lubricant-condition trending and points users to equipment and lubricant criteria for limits; it does not supply a universal oxidation cutoff.
Decide what to investigate next
In this example, ask the laboratory to examine B's change from 0.20 to 0.25, including whether it is analytically meaningful. Review service conditions and supporting results such as viscosity and acid number. Do not dismiss the rise simply because the refill explains the earlier step.
If the apparent jump instead comes from a changed method or unsuitable reference, correct the comparison and keep the original reports. If no suitable unused sample exists, say so. An assumed zero makes the chart neater but the decision weaker.
The oil change belongs on the trend as an event. The question after it is how the new fill is changing from its own start.
References
- Thermo Fisher Scientific, FT-IR Analysis of Used Lubricating Oils: unused-oil spectra and reference selection, page 2.
- Spectro Scientific, Guide to Measuring Oil Chemistry: oxidation chemistry and infrared measurement.
- ASTM D7414-22: public scope for oxidation trending, reporting conventions and lubricant applicability.
- ASTM D7418-23: public scope for FTIR instrument setup and operation.
- ASTM D7669-20: public scope for practical lubricant-condition trend analysis.